61
5. On the Main Menu screen, Select
SERVICE
, and press
[OK]
. Enter Password:
[UP] [DOWN] [LEFT] [RIGHT] [OK]
.
After entering the Servicer password,
select
BLOWER SPEED
, and press
[OK]
.
6. You will be prompted for an Admin Password. Enter Password:
[UP] [RIGHT]
[DOWN] [START] [OK]
.
7. With the
HUMIDIFY SPEED
field highlighted, press
[RIGHT]
until the speed is
95
. Press
[OK]
.
8. Go to the Main Menu screen, Select
RUN
and press
[OK]
.
Select
CLEANING
and press
[OK]
. Do
NOT
press
[START]
yet.
9. Open the Main Door slightly. Turn the shopvac or similar vacuum on.
10. Press
[START]
.
11.
Allow the Fuming Chamber’s blowers to run for 1 minute. Press
[STOP]
.
12. Remove the vacuum tube, and close both the Main Door and CA Chamber (Hot
Plate) Doors.
13. Go to the Main Menu screen, Select
SERVICE
, and press
[OK]
. Enter
Password:
[UP] [DOWN] [LEFT] [RIGHT] [OK]
.
After entering the Servicer
password, select
BLOWER SPEED
, and press
[OK]
.
14. You will be prompted for an Admin Password. Enter Password:
[UP] [RIGHT]
[DOWN] [START] [OK]
.
15. With the
HUMIDIFY SPEED
field highlighted, press
[LEFT]
until the speed is
20
.
Press
[OK]
.
Summary of Contents for Capture BT 31700-00
Page 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Page 55: ...55 Figure 7 6...
Page 81: ...81 Wiring Diagrams 100 120V...
Page 82: ...82 208 230V...
Page 87: ...87 Chamber will not achieve humidity...
Page 88: ...88 Chamber does not exhaust CA fumes...
Page 89: ...89 Door alarm active when door s closed...
Page 90: ...90 Hot Plate will not achieve temperature set point...
Page 91: ...91 White CA residue builds on inside of Chamber...