64
Moving the Chamber
Once installed, the CApture BT Fuming Chamber should not be moved or tipped without
first preparing the chamber. If the chamber is installed on the accessory stand, it can
be rolled without any preparation.
Do not roll the chamber on the accessory stand over rough surfaces, door thresholds, or
on uneven surfaces of more than 5 degrees of inclination.
Always remove all
evidence and CA tin before moving the chamber.
To prepare the chamber for its move, perform the following tasks:
1. Remove all Hang Bars, and any accessory shelves or other accessories
installed inside the Main Chamber. Remove the Perforated Floor from inside
the Main Chamber.
2. Open the Water Bottle Door, locate the Valve Toggle (see Fig. 7-13) on the
back wall, just to the left of the Water Bottle. Next, flip the Valve Toggle “up”
which reverses water flow back into the Water Bottle.
3. Make sure the unit is plugged in, and Main Power Switch is ON. Locate the
Momentary Switch (see Fig. 7-13) on the top panel inside Water Bottle
Compartment. Press and hold this switch for approximately 15 seconds.
When the switch is pressed, you should hear the Water Pump turn on, and
see the water level in the Bottle
rising
. Release the Switch, Water Pump
should continue to run on its own, if not, hold Switch for another 5 seconds.
4. When air bubbles are seen coming from hose inside Water Bottle, turn the
Main Power Switch OFF. Unplug the unit’s power cord. Empty the Water
Bottle of all water. Flip the Valve Toggle back “down” to the NORMAL FILL
position. Replace Water Bottle.
Figure 7-13
MOMENTARY
SWITCH
VALVE
TOGGLE
WATER
BOTTLE
Summary of Contents for Capture BT 31700-00
Page 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Page 55: ...55 Figure 7 6...
Page 81: ...81 Wiring Diagrams 100 120V...
Page 82: ...82 208 230V...
Page 87: ...87 Chamber will not achieve humidity...
Page 88: ...88 Chamber does not exhaust CA fumes...
Page 89: ...89 Door alarm active when door s closed...
Page 90: ...90 Hot Plate will not achieve temperature set point...
Page 91: ...91 White CA residue builds on inside of Chamber...