59
11. Reinstall the Humidity Sensor in to the housing in the Water Bottle Compartment.
12. Once the Sensor is in place, the O-Ring should rest against the housing, and the
Humidity Sensor Bracket will be loose. Position the three (3) large cutouts on the
Bracket over the three (3) studs on the housing. Slide the Bracket toward the O-
Ring and compress the O-Ring slightly by pushing the Bracket towards the
housing. While compressing the O-Ring, rotate the Bracket clockwise until it
locks in place.
13. Connect Harness Connector at the end of the Humidity Sensor to the thru-wall
connector mate as shown in Fig. 7-8.
14. After installing a new Humidity Sensor, the software must be recalibrated to
accurately display relative humidity. See
Recalibrate Humidity Sensor (Annually)
previously in this section to calibrate the new Humidity Sensor.