16
o
Always dispose of disinfecting solutions in accordance with local and national
laws.
o
DO NOT allow disinfectants with high concentrations of free chlorine to
contact the stainless steel components of the chamber for a long period of
time. Free chlorine will corrode stainless steel after extended contact.
Never use the chamber as a chemical storage cabinet.
When handling evidence, or working in the chamber, always wear proper
personal protective equipment (PPE). Proper PPE should provide skin, eye, and
breathing protection from CA fumes and contact with liquid CA. Never wear
cotton gloves when working with CA.
This product only uses Carbon Filtration to remove vapors from the exhaust air.
Carbon Filtration only removes vapors, biohazards may pass through the filter.
The Main Carbon Filter will gradually accumulate Cyanoacrylate (CA) from the
fuming cycles run in the chamber. The rate and amount of accumulation will
depend upon the amount of CA used, the frequency of Fuming Cycles, and
factors described in section
Carbon Filter Life
on the following page.
ALWAYS
replace the Main Carbon Filter when prompted to do so by the Display
with
a new Filter from Labconco. NEVER reset the Carbon Filter Life Gauge without
actually replacing the Filter with a new filter from Labconco.
Do not operate the chamber without all of the appropriate filters in place. If
Cyanoacrylate is detected coming out of the Main Carbon Filter, it needs to be
replaced, even if the Display has not yet prompted the user to change the Filter.
When handling used filters or pre-filters, always wear appropriate personal
protective equipment (PPE).
Never tip the chamber after installation. Water in the humidifier could spill
–
damaging internal components. See
Section 7: Moving the Chamber
for
instructions on removing the water from the Humidifier Basin before tipping or
moving the chamber, or contact Labconco at 800-821-5525 or 816-333-8811 for
further information.
Summary of Contents for Capture BT 31700-00
Page 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Page 55: ...55 Figure 7 6...
Page 81: ...81 Wiring Diagrams 100 120V...
Page 82: ...82 208 230V...
Page 87: ...87 Chamber will not achieve humidity...
Page 88: ...88 Chamber does not exhaust CA fumes...
Page 89: ...89 Door alarm active when door s closed...
Page 90: ...90 Hot Plate will not achieve temperature set point...
Page 91: ...91 White CA residue builds on inside of Chamber...