58
4. Disconnect the Harness Connector at the end of the Humidity Sensor from the
thru-wall connector mate. To disconnect, press down on the small lever tab on
the connector and pull the Harness Connector towards you.
5. Rotate the Humidity Sensor Plate counter-clockwise approximately 30 degrees
until it pops free from the three retainer fasteners.
6. Slide the Humidity Sensor body straight out. There is an O-Ring on the Humidity
Sensor, this will come with the Humidity Sensor.
7. Roll the O-Ring off the Humidity Sensor, and then remove the Humidity Sensor
Plate. The Humidity Sensor Plate will be the only item saved when installing the
new Humidity Sensor. See Fig. 7-9 below.
8. Identify the components inside the new Humidity Sensor Kit (P/N 3179230). The
kit includes a new Sensor, Sensor Filter, and O-Ring.
9. Peel the protective backing off the new Humidity Sensor Filter, and apply the
filter onto the sensor, taking care to position the rectangular opening in the filter’s
adhesive over the slots in the new Humidity Sensor. Wrap the Filter completely
around the sensor and press it tight in all areas.
10. Slide the original Humidity Sensor Bracket onto the new Sensor as shown in
Step 5 of Fig. 7-9. Roll the new O-Ring over the end of the sensor and position it
approximately as shown in Step 5 of Fig. 7-9
OLD
HUMIDITY
SENSOR
HUMIDITY
SENSOR
PLATE
NEW
HUMIDITY
SENSOR
FILTER
NEW
HUMIDITY
SENSOR
NEW
O-RING
OLD
O-RING
Figure 7-9
1
2
3
4
5
Summary of Contents for Capture BT 31700-00
Page 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Page 55: ...55 Figure 7 6...
Page 81: ...81 Wiring Diagrams 100 120V...
Page 82: ...82 208 230V...
Page 87: ...87 Chamber will not achieve humidity...
Page 88: ...88 Chamber does not exhaust CA fumes...
Page 89: ...89 Door alarm active when door s closed...
Page 90: ...90 Hot Plate will not achieve temperature set point...
Page 91: ...91 White CA residue builds on inside of Chamber...