76
Heater Wire from Removable WAGO Wall Nut Connector behind CA Chamber,
disconnect Red Angle Terminal from Item 32. Feed wires through grommets in back
side wall of CA Chamber (may have to peel away RTV Silicone to push wires
through grommets). Reseal wires w/ Silicone.
32. Temp Controller
– P/N 1291501
Remove Items 33 & then 31 as described. Remove two (2) wires from Temp
Controller. Remove two (2) nuts holding Temp Controller to Heater Plate.
33. Tin Holder
– P/N 3172100
Remove (2) Tin Holder Screws, lift up & angle to pull into Main Chamber.
Summary of Contents for Capture BT 31700-00
Page 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Page 55: ...55 Figure 7 6...
Page 81: ...81 Wiring Diagrams 100 120V...
Page 82: ...82 208 230V...
Page 87: ...87 Chamber will not achieve humidity...
Page 88: ...88 Chamber does not exhaust CA fumes...
Page 89: ...89 Door alarm active when door s closed...
Page 90: ...90 Hot Plate will not achieve temperature set point...
Page 91: ...91 White CA residue builds on inside of Chamber...