ELECTRICAL CONNECTIONS
4
49
OPTIFLEX 8200 C/F/S
www.krohne.com
06/2018 - 4006448901 - MA OPTIFLEX 8200 R01 en
Terminals for electrical installation (two current outputs)
Figure 4-2: Terminals for electrical installation (two current outputs)
1
Output 1: Terminals
2
Output 2: Terminals
3
Location of the external grounding terminal (at the bottom of the converter)
4
Grounding terminal in the housing (if the electrical cable is shielded)
CAUTION!
•
Use the applicable electrical cables with the cable glands.
•
Make sure that the current is not more than 4 A or that there is 4 A-rated fuse in the
electrical circuit that energizes the device.
•
Make sure that the polarity of the power supply is correct. If the polarity is incorrect, you will
not cause damage to the device but the device will not operate.