SERVICE
7
139
OPTIFLEX 8200 C/F/S
www.krohne.com
06/2018 - 4006448901 - MA OPTIFLEX 8200 R01 en
Procedure: how to disassemble the remote version
•
Remove the housing base (8) and the parts attached to the housing base. Remove the PCB (12)
PCB (12)
PCB (12)
PCB (12).
•
Remove the probe electronics housing (32) and the parts attached to the probe electronics
housing. Remove the PCB (31)
PCB (31)
PCB (31)
PCB (31).
•
If the device has an Ex d approval or the second output option, remove the Ex d housing (23)
and the parts contained in the Ex d housing. Remove the PCB (30)
PCB (30)
PCB (30)
PCB (30) from the Ex d housing.
•
Remove the housing wall support (28) and the parts attached to the housing wall support.
•
If the device has the display screen option, remove the housing cover (21) and the parts
attached to the housing cover. Remove the PCB (20)
PCB (20)
PCB (20)
PCB (20) and the attached display screen
display screen
display screen
display screen.
•
Remove the PCB (17)
PCB (17)
PCB (17)
PCB (17) from the "T" housing (16).
•
If the device has an Ex d approval or the second output option, remove the Ex d housing (23)
and the parts contained in the Ex d housing. Remove the PCB (24)
PCB (24)
PCB (24)
PCB (24) from the Ex d housing.
Figure 7-3: Parts of the device (refer to "Parts list" table)