GG-09001 Edition 1.3
-
58
-
13-3
Adjustment of the black point glass unit
①
Remove the covers according to “5-2. Removing optical head cover”.
②
Remove cables from motor and sensor PCB, and then remove inner cover by loosening 2
screws.
③
Turn on the power switch and rotate “observation light intensity control knob” to maximum,
after connecting cables with motor and sensor PCB.
2
3
④
Put the model eye on chin rest, and then move the optical head unit to focus the illumination
light onto black surface at the model eye.
⑤
Insert the minus compensation lens by pressing “Diopter switch (-)”.
⑥
Rotate focusing knob to focus on the center spot in finder view.
4
5
6
⑦
Loosen two screws for fixing the black point glass unit.
⑧
Adjust the black point lens unit to hide the center spot, and then fix it by two screws.
⑨
Install inner cover and outer cover.
(
(
(
(
Check the result by taking photos at each photographic modes.
)
)
)
)
7
8
Cables
Screws
Black point glass unit
Summary of Contents for K9L48B
Page 1: ...GG 09001 Edition 1 3 Edition 1 3 2015 4 Model Name K9L48B...
Page 4: ...GG 09001 Edition 1 3 1 Names and Functions of Components...
Page 5: ...GG 09001 Edition 1 3...
Page 8: ...GG 09001 Edition 1 3 Principals of Fundus Photography...
Page 9: ...GG 09001 Edition 1 3 3 Block Diagram...
Page 23: ...GG 09001 Edition 1 3 Optical Head Unit Fan Assembly Drawing...