GG-09001 Edition 1.3
-
25
-
6-8 Replacing upper mount PCB
①
Undo 2 screws and remove the dust-proof cover.
②
Remove the tying band shown in the picture below.
③
In order to allow cables to be easily pulled out, loosen the screws (to loosen the sheet metal
that holds the cables behind it).
④
Undo 2 screws from the Polaroid/video mount contact PCB.
⑤
Disconnect the connector from the Polaroid/video mount contact PCB.
⑥
To install the PCB, follow the removing steps in the reverse order. Make sure that all cables
are placed behind the sheet metal, tighten the screws shown in the step
③
to hold the cables,
and tie the cables as shown in the step
②
. Excessively loose cables may be captured into a
photograph, interrupt the mirror movement, or damage the mirror.
※
※
※
※
Carefully proceed to the work steps since this segment contains fragile components
including lenses and mirrors.