GG-09001 Edition 1.3
-
56
-
14. Additional information
13-1. How to step for incorrect information of left or right
13-1-1. How to check
①
Put the model eye on the chin rest. Make the direction of the model eye to face the front of
the optical head unit.
②
Unlock the lock lever. Move the base to the direction of the operator, and align the left and
right of the base and power supply unit. Then lock the lock lever.
③
Check that the position of the swing arm is center. (Recommend to the 6
th
step, if the swing
arm is already center.)
④
Unlock the horizontal lock lever.
⑤
Align the swing arm to center, and lock the horizontal lock lever.
2
3, 4
5
⑥
Make the optical head unit to horizontal by operating tilting handle. It sounds click when the
optical head unit made to horizontal.
⑦
Adjust the height level between aperture of the model eye and objective lens to equal by
operating “Camera height adjusting handle” on the monitor like below figure.6.
⑧
Check that aperture of the eye model is within between two vertical lines on the monitor.
⑨
It needs adjustment, if the aperture was not within between two vertical lines.
6
7, 8
9