TOS9200/9201
Basic Operations 3-47
If the key is not locked, the test voltage can be changed during the test using the
rotary knob.
The cursor moves to the 2nd decimal place in ACW and IR, and to the 1st decimal
place in DCW.
• In an AC withstanding voltage test with the output voltage range set to AUTO, if
the test starts with the test voltage set to 2.60 kV or less, the voltage cannot be
reset to 2.60 kV or more during the test.
• If the test voltage is changed when the measured current is close to the upper cur-
rent, the test may end with a FAIL judgement.
Screens can be switched during the test using the F5 key (DISP). To suspend the
test, press the STOP switch.
Example of an AC withstanding voltage test (DISP1)
Example of an AC withstanding voltage test (DISP2)
Example of a DC withstanding voltage test (DISP1)
NOTE
DISP
2.50
kV
10
A
3.0
s
ACW
ACW
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
OFF
TEST
NORM
60Hz
Absolute value of measured current
DISP
TEST
2.50
kV
1.4
s
12
A
NORM
ACW
ACW
REAL
11 A
IMAG
+J 1 A
Absolute value of measured current
Real-value portion of the measured current
Imaginary-value portion of the measured current
DISP
1.00
kV
10
A
3.0
s
DCW
DCW
LOWER
OFF
TIMER
OFF
TEST
NORM
Measured current
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......