TOS9200/9201
Controlling the TOS6200 5-1
Chapter 5
Controlling the TOS6200
This chapter describes the procedure for using the TOS9200/TOS9201 to control
Kikusui’s earth continuity tester TOS6200 via the RS-232C interface.
This control function is used to recall and execute the test conditions stored in the
TOS6200’s memory by executing the TOS9200/TOS9201 program.
For details on the RS-232C interface, see the Operation Manual for the GPIB and
RS-232C Interfaces separately available.
• The GPIB interface cannot be used to control the TOS6200 with the TOS9200/
TOS9201.
During control of the TOS6200, the TOS9200/TOS9201 cannot be controlled
using the GPIB interface.
5.1
Pre-Control Preparation
5.1.1
Connection and startup procedure
1.
Turn off the POWER switch on the TOS9200/TOS9201 and the
TOS6200.
2.
Using an RS-232C cable, connect the RS-232C connectors on the rear
panel of the TOS9200/TOS9201 and the TOS6200 (use a cross cable
of the 9-pin D-SUB female-female AT type as the RS-232C cable).
3.
Turn on the POWER switch on the TOS9200/TOS9201 and the
TOS6200.
5.1.2
Settings on the TOS6200
Interface settings
On the TOS6200, press the SHIFT + SYSTEM /I/F keys. The LED on the SYSTEM
/I/F key will then light up, and the Interface Settings screen (INTERFACE) will
appear.
On this screen, make the settings speci
fi
ed below.
SPEED (baud rate)
: 19
200 bps
DATA (data length)
: 8 bits
PARITY
: NONE
STOP (stop bit)
: 2 bits
Upon completion of the above settings, turn the POWER switch off and then on
again.
For details on the setting procedure, see the Operation Manual for the TOS6200.
NOTE
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......