5-6 Controlling the TOS6200
TOS9200/9201
• During the execution of a program, no buzzer may sound even if the TOS6200
makes a PASS judgement. This does not indicate any abnormality.
Displaying a communication error
In the following cases, “COM ERR”
fl
ashes and the PROTECTION LED lights up
to indicate a communication error:
• The POWER switch on the TOS6200 is turned off.
• The interface settings for the TOS6200 are not made as instructed in "5.1.2
• The RS-232C cable is disconnected or broken.
• At the start of a test, the TOS6200 is not in the READY status.
To cancel a communication error, press the STOP switch. Eliminate the cause of the
communication error, and restart the test.
PROTECTION
During the execution of the TOS9200/9201 program, if the PROTECTION function
activates on the TOS6200, “TOS6200”
fl
ashes at the top right of the screen on the
TOS9200/TOS9201, and the PROTECTION LED lights up.
To cancel the PROTECTION function on both the TOS9200/9201 and the
TOS6200, press the STOP switch on the TOS9200/TOS9201.
A test using the TOS6200 may not start, depending on the cause of PROTECTION.
To restart the test, eliminate the cause of PROTECTION and con
fi
rm that the tester
is in the READY status.
5.4
Canceling the TOS6200 Control Mode
The TOS9200/9201 is ready to control the TOS6200 via the RS-232C interface, so
the POWER switch must
fi
rst be turned off. To cancel the control mode, take the fol-
lowing steps:
1.
On the TOS9200/9201 and the TOS6200, turn off the POWER switch.
2.
On the rear panel of the TOS9200/9201 and the TOS6200, disconnect
the RS-232C cable from the RS-232C connector.
To use either the TOS9200/9201 or the TOS6200 alone, turn on the POWER switch.
NOTE
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......