TOS9200/9201
Specifications 8-3
Item
Speci
fi
cations
TOS9200
TOS9201
Ammeter *3
Measurement range
0.00 mA to 110 mA AC
0.00 mA to 110 mA AC
/0.00 mA to 11 mA DC
Display
i = measured current
i < 1 mA
1 mA
≤
i < 10 mA
10 mA
≤
i < 100 mA
100 mA
≤
i
μ
A
.
mA
. mA
mA
Accuracy
±(3% of the r 20
μ
A)
[after the offset cancel function is activated, if the scanner is mounted]
Response
Mean-value responsive / root-mean-square value display (response time of 200 ms)
Hold function
The measured current at the end of the test is held during the PASS period.
Offset cancel function
The current
fl
owing to the insulation resistor between the output cables and the stray
capacity is cancelled up to 100
μ
A/kV (in AC withstanding voltage testing only).
Calibration
Performs calibration using the root-mean-square value of a sine wave with a pure
resistive load
Selection of LOW/GUARD
for the GND (*4)
Selection permitted for current measurement between the mode for the GND point
connected to the LOW terminal, and the mode using guard.
LOW
Connects the GND point to the LOW terminal. Measures the current
fl
owing to the
LOW terminal (chassis) (for normal operation).
GUARD
Sets the GND point as guard. Measures the current
fl
owing to the LOW terminal,
but does not measure the current
fl
owing to the chassis (for high-sensitivity, high-
accuracy measurements).
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......