TOS9200/9201
Basic Operations 3-59
2.
Using the F2 key (CONTRAST) or the keys, move the cursor to
CONTRAST.
3.
Using the rotary knob, adjust the contrast level.
The contrast level can be set using the SHIFT + keys, regardless of the screen
displayed.
3.10.3 SYSTEM3
Setting the STATUS SIGNAL OUTPUT
Make settings to specify the timing of the output of a DC voltage of 24 V from the
STATUS OUT connector on the rear panel.
To make settings for the STATUS SIGNAL OUTPUT terminal, use the function
keys or the rotary knob.
Each time a function key or the SHIFT + function keys are pressed, the item shown
on the key is turned on/off.
The mark appears to the left of the selected item. Using the rotary knob, the items
shown with the mark on the left can be set.
Turn the rotary knob clockwise
: ON
Turn the rotary knob counterclockwise : OFF
Items set to ON are highlighted.
H.V ON
When H.V ON is set to ON, DC 24 V is output during the test, while a voltage
remains between the output terminals, or during automatic testing.
TEST
When TEST is set to ON, DC 24 V is output during the test after the test voltage
reaches the preset value. That is, no voltage is output during the RISE time or FALL
time.
H.V ON
TEST
PASS
READY
SYSTEM 3
SYSTEM 3
STATUS SIGNAL
OUTPUT :
F1 F2 F3 F4 F5
OFF ON
H.V ON
TEST
PASS
READY
OFF ON
U FAIL
L FAIL
C FAIL
PROTECT
POWER ON
SYSTEM 3
SYSTEM 3
STATUS SIGNAL
OUTPUT :
F1 F2 F3 F4 F5
OFF ON
H.V ON
TEST
PASS
READY
OFF ON
U FAIL
L FAIL
C FAIL
PROTECT
POWER ON
UPPER
FAIL
LOWER
FAIL
CONTACT
FAIL
PROTEC-
TION
POWER
ON
SHIFT
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......