3-46
Basic Operations
TOS9200/9201
3.8
Starting and Ending a Test
•
The test cannot be started with in
v
alid settings or in the protection status. F
or
information on in
v
alid settings and the protection status, see
.
•
The test cannot be started while the ST
OP switch is pressed. (The test also cannot
be started when the ST
OP signal of remote control is acti
v
e.)
•
If the DOUBLE
A
CTION is on, press the ST
AR
T switch within approximately 0.5
seconds after pressing the ST
OP switch, as the test will not start otherwise. F
or
details on DOUBLE
A
CTION, see
•
If the MOMENT
AR
Y function is ON, the test runs only while the ST
AR
T switch
is pressed. F
or details on MOMENT
AR
Y
, see
3.8.1
Starting a test
To start a test, press the START switch when the tester is in the READY status on
the Test-Condition Settings screen. When the tester is in the READY status,
“READY” lights up at the top right of the LCD.
Example of an AC withstanding voltage test (READY)
As soon as the test starts, the LCD displays the following:
During the test, “TEST” is displayed at the top right of the LCD, and the TEST LED
and DANGER lamp light up on the indicator. (During a voltage rise, “TEST”
fl
ashes at the top right of the LCD.)
Example of an AC withstanding voltage test (TEST)
The time is displayed differently depending on whether the timer function is on or
off.
The timer is ON
: The remaining test time is displayed.
The timer is OFF : The elapsed test time is displayed.
(When 999 seconds have elapsed, “999”
fl
ashes.)
NOTE
VOLT LOWER UPPER
TIMER
2.50
kV
10.0
mA
3.0
s
ACW1
ACW1
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
OFF
READY
UPPER
60Hz
DISP
2.50
kV
11
A
3.0
s
ACW
ACW
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
OFF
TEST
NORM
60Hz
Summary of Contents for TOS9200