3-10 Basic Operations
TOS9200/9201
with a PASS judgement. The timer function can be set to ON/OFF using the SHIFT
+ F5 keys, regardless of the location of the cursor. Each time the SHIFT + F5 keys
are pressed, ON and OFF alternate.
The keys can also be used to move the cursor to ON/OFF of the timer
function. When the cursor is at ON/OFF of the timer function, the rotary knob can
be used to make settings.
Turn the rotary knob clockwise
: ON
Turn the rotary knob counterclockwise : OFF
Setting the test time (TEST TIME)
•
If the upper current is set above 50 mA, the protective circuit may be acti-
vated. To avoid this, set the output time to 30 minutes or less and provide a
pause longer than the output time.
The time for which a preset voltage is applied to the DUT can be set to 0.3 s through
999 s (at a resolution of 0.1 s for the 0.3 s to 99.9 s range and 1 s for the 100 s to 999
s range).
When the cursor is at the test time, the rotary knob can be used to make settings.
1.
To move the cursor to the timer, press the F5 key (TIMER). (The
keys can also be used.)
2.
Using the keys, move the cursor to the target digit.
3.
Using the rotary knob, set the test time.
CAUTION
VOLT LOWER UPPER
TIMER
1.00
kV
10.0
mA
3.0
s
ACW1
ACW1
FREQ
LOWER
OFF
OFFSET
OFF
TIMER
ON
READY
UPPER
50Hz
F1 F2 F3 F4 F5
SHIFT
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......