6-8
PAN-A
Output current in full scale
14. Set the output current to the rated output current.
For local control, turn the CURRENT control knob fully clock-
wise. For remote control, set the control signal to 10.5 V or 10
k
Ω
.
15. Using variable resistor "7", adjust the output current to 105
% of the rated output current.
• The maximum output current of the unit is speci
fi
ed
as 105 % of the rated output current. It is possible to
adjust to a level higher than 105 % using variable
resistor "7", but this may damage the unit.
• For remote control using external voltage, a 10 V control sig-
nal may be selected and adjusted to 100 % of the rated output
current of the unit.
Output current display in full scale
16. Adjust with variable resistor "8" so that the output current
indication of the unit equals the current value obtained from
the indication on the external digital voltmeter (DVM) and
shunt resistor, in a rated-current output condition.
Fig.6-6
Current system adjustment variable resistor
CAUTION
NOTE
2
4
5
6
7
8
S1
S2
S3
S4
S5
OVP
11
Front control panel
1
9
10
3
Output current offset
Output current display in full scale
Output current in full scale
Never touch
"1", "3", "9" or "10".