Turning on the power
Before turning on the POWER switch, always check the status of the
OUTPUT switch. Push in the OUTPUT switch to turn it on, and
release it to turn it off.
• If the POWER switch is turned on while the OUTPUT
switch remains on, a preset voltage or current is sup-
plied to the load.
• If the POWER switch is turned on while the OUTPUT switch
remains on, AC ripple components may be superimposed at
startup depending on the operation speed of the phase control
circuit and the status of the load. Even when the OUTPUT
switch is turned on after a power input, AC ripple components
may be superimposed on the waveform at startup. This may
occur if the input line has an excessively high impedance or if
the input voltage is too low.
• When the remote sensing function is not being used, be sure
to turn off the SENS switch. Turning on the SENS switch can
cause difference between output voltage and display.
CAUTION