4-34
PAN-A
7. Turn on the OUTPUT switch of the master unit.
"CV" lights up on the master unit control panel, indicating that
the unit is in constant voltage operation mode.
The same output voltage and current displayed on the master unit
are also displayed on all the slave units. This means that the total
current of all the units is applied to the load.
Finishing the master-slave-control parallel operation
• To
fi
nish master-slave-control parallel operation,
always observe the following procedure. If the
POWER switch of the master unit is turned off while
parallel operations are still being carried out, the
slave units may output their maximum output voltage.
1. Turn off the OUTPUT switch of the master unit.
2. Turn off the OUTPUT switch of each slave unit.
3. Turn off the POWER switch of each slave unit.
4. Turn off the POWER switch of the master unit.