1
7. Connect an electrolytic capacitor (C) with a capacity of a
few hundreds of
μ
F to a few tens of thousands of
μ
F to the
load terminals as necessary.
• Use a capacitor (C) whose withstand voltage is 120
% or more of the unit's rated voltage.
• If the length of wiring to a load extends to 3 m or more, the
phase shift caused by the inductance and capacitance of the
wiring becomes non-negligible, thereby causing oscillation.
In such cases, the capacitor (C) prevents oscillation.
8. Re-check that the connections and the switch settings are
correct.
• When using shielded wires for connection, protect
the uncovered section of the shielded wire by using
insulation tubes with a withstand voltage greater than
the isolation voltage of the unit.