PAN-A
7-1
Chap
.7
Specifications
1
Chapter 7
Speci
fi
cations
Unless otherwise speci
fi
ed, the speci
fi
cations of the unit are based on
the following conditions.
• The load is a pure resistance.
• The - (neg.) output terminal is connected to the (
⊥
) chassis ground
terminal with the supplied shorting bar.
• The unit should be used after 30 minutes warming-up time (with
current
fl
owing).
TYP value, standard value and theoretical value do not guarantee per-
formance. They should be referred to as target values only.
Common Speci
fi
cations
Common Speci
fi
cations
Input
Input voltage and fre-
quency
100 VAC±10 %, 50/60 Hz, 1-phase, Crest factor: 1.2 to 1.41
(110 V, 120 V, 200 V, 220 V, 230 V and 240 V AC input are fac-
tory option.
Constant voltage operation
indication
C.V, green LED indication
Constant current operation
indication
C.C, red LED indication
Range of operation tem-
perature and humidity
0
˚
C to 40
˚
C (32
˚
F to 104
˚
F)
10 % to 90 % RH (no dew condensation allowed)
Range of storage tempera-
ture and humidity
-10
˚
C to 60
˚
C (14
˚
F to 140
˚
F)
0 % to 90 % RH max. (no dew condensation allowed)
Cooling system
Forced air cooling with fan
Output polarity
Positive or negative grounding possible
Withstanding voltage
Across input terminals
and output terminals
Should withstand 1500 V AC, 1 min. with no abnormalities
Across input terminals
and chassis
Remote sensing
Possible (compensation one way approx. 0.6 V max.)
Master-slave-control paral-
lel operation
Possible
*1
For 230 V and 240 V AC input requirement, maximum input voltage is limited to 250 V AC.