Symptom 4: No output is generated even when the OUT-
PUT switch is turned on.
Check item
Cause and remedy
Check to see if both the
"CV" and "CC" LEDs are
off.
• The VOLTAGE and CURRENT control
knobs have been turned fully counterclock-
wise. Both LEDs may not light up,
depending on the offset setting conditions.
This is not a form of equipment failure. Set
the required output using the VOLTAGE
and CURRENT control knobs.
Check that the VOLTAGE
or CURRENT control
knob has not been turned
fully counterclockwise.
• Set the required output using the VOLTAGE
or CURRENT control knob.
Check to see if the "OUT-
PUT ON" LED is off.
• The external contact points have been
closed in output ON/OFF control using
external contact points. See "4.3.5 Output
ON/OFF control".
Check that the switches
S1, S3, and S5 on the con-
trol panel have not been
set to on (upper position).
• For control from the front panel, control
switches S1, S3, and S5 should all be set to
off (lower position). See "4.3 Analog
remote control".