Applied Oper
ation
1
Connection and setup procedure
1. Turn off the OUTPUT switch of each unit.
2. Turn off the POWER switch of each unit.
3. Determine the master unit.
• If the PAN-A and PAN series are used together, assign the
PAN-A series power supply as the master unit.
4. Preset the OVP (overvoltage protection) trip points for the
master and slave units.
• Preset the OVP trip points for the slave units to a value
slightly higher than that used for the master unit. However,
do not preset them to a value exceeding 105 % of the rated
voltage.
5. Set the control switch S5 of the master unit and slave units
to the positions shown in Fig.4-17.
Fig.4-17
Control switch presetting for master-slave-
control parallel operation
6. Connect the master unit and slave unit as shown in Fig.4-
Fig.4-18 shows an example of how the rear output terminals are
connected in parallel for two slave units
For the PAN16-50A user, see Fig.4-19.
To combine the PAN-A and PAN series, see Fig.4-20.
NOTE
NOTE
1
2
3
4
5
6
7
8
9
10
S1
S2
S3
S4
S5
OVP
11
1
2
3
4
5
6
7
8
9
10
S1
S2
S3
S4
S5
OVP
11
Slave unit Front control panel
Master unit Front control panel
Set S5 of the master
unit downward
Set S5 of the slave
units upward.