85
Command Reference
Sense Subsystem
Average Subsection
Average Type
[:SENSe]:AVERage:TYPE LOGPower|POWer|VOLTage
[:SENSe]:AVERage:TYPE?
Toggle the average type between Log power, power and
voltage.
Example:
:AVERage:TYPE POW
Key access:
BW > Average Type
Average Number On/Off
[:SENSe]:AVERage:TRACe[1]|2|3|4:COUNt <integer>
[:SENSe]:AVERage:TRACe[1]|2|3|4:COUNt?
Specifies the number of measurements that are combined.
Range:
1 to 8192
[:SENSe]:AVERage:TRACe[1]|2|3|4[:STATe]
OFF|ON|0|1
[:SENSe]:AVERage:TRACe[1]|2|3|4[:STATe]?
This command toggles averaging off and on.
Example:
:AVERage:TRACe1:COUNt 20
Key access:
Trace > More > Average
Average Duration On/Off
[:SENSe]:AVERage:TRACe[1]|2|3|4:DURation <value>
[:SENSe]:AVERage:TRACe[1]|2|3|4:DURation?
Sets the duration of average process.
[:SENSe]:AVERage:TRACe[1]|2|3|4:DURation:STATe
OFF|ON|0|1
[:SENSe]:AVERage:TRACe[1]|2|3|4:DURation:STATe?
This command toggles the averaging duration off and on.
Example:
:AVERage:TRACe1:DURation 2S
Key access:
Trace > More > Average Duration On/Off