Command Reference
Calculate Subsystem
94
Marker Subsection
Markers All Off on All Traces
:CALCulate:MARKer:AOFF
Turns off all markers on all the traces.
Key access:
Marker
> All Off
Continuous Peaking Marker
:CALCulate:MARKer[1]|2|3|4|5|6:CPEak[:STATe]
OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4|5|6:CPEak[:STATe]?
Toggles the continuous peak search function between on and
off.
Example:
:CALCulate:MARKer1:CPEak ON
Key access:
Shift > Peak > More > Continuous PK
Frequency Counter Marker
:CALCulate:MARKer[1]|2|3|4|5|6:FCOunt[:STATe]
OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4|5|6:FCOunt[:STATe]?
Turns on/off the marker frequency counter. To query the
frequency counter, use
:CALCulate:MARKer[1]:FCOunt:X?
If the specified
marker number is not the active marker, it becomes the
active marker. If the specified marker number is not on, it is
turned on and becomes the active marker. A 1 is returned
only if marker count is on and the selected number is the
active marker.
*RST:
Off
Remarks:
If a frequency count x value is generated when the
frequency count state is off, then 0 is returned.
Example:
:CALCulate:MARKer1:FCOunt 1
Key access:
Marker > Function > Counter