135
Command Reference
Reflection Measurement Option Subsystem
Select A Cable in Reflection Measurement
[:SENSe]:CAT:CABLe:LOAD <string>
Use this command to select a cable from built- in cable list.
Example:
:CAT:CABLe:LOAD "5088"
Key access:
Meas > More 1 of 2 > Cable Specification > Select Cable
Reflection Measurement Cable Attenuation
[:SENSe]:CAT:CABLe:ATTenuation <value>
[:SENSe]:CAT:CABLe:ATTenuation?
This command defines the attenuation value for a custom
cable.
Range:
0 to 5 dB/m, or 0 to 1.524 dB/ft
Example:
:CAT:CABLe:ATTenuation 1.2
Key access:
Meas > More 1 of 2 > Cable Specification > Cable Type > Custom, then
Cable Atten
Reflection Measurement Cable Velocity Factor
[:SENSe]:CAT:CABLe:VELFactor <value>
[:SENSe]:CAT:CABLe:VELFactor?
This command defines the velocity factor for a custom cable.
Range:
1 to 100
Example:
:CAT:CABLe:VELFactor 89.2
Key access:
Meas > More 1 of 2 > Cable Specification > Cable Type > Custom, then
Vel Factor