Command Reference
Channel Scanner Option Subsystem
160
Display Threshold State
:CHSCanner:DISPlay:THReshold [:STATe] OFF|ON|0|1
:CHSCanner:DISPlay:THReshold [:STATe]?
This command toggles display threshold state. Then the
display threshold is turned on, only the channels with higher
channel power than the threshold value will be displayed on
the screen.
Remark:
This parameter is only available for Table (Bar) display
mode.
Example:
:CHSCanner:DISPlay:THREShold 1
Key access:
Shift > Disp > Disp Mode > Threshold
Display Threshold Value
:CHSCanner:DISPlay:THReshold:VALue <value>
:CHSCanner:DISPlay:THReshold:VALue?
This command defines the display threshold value.
Example:
:CHSCanner:DISPlay:THReshold:VALue -70
Key access:
Shift > Disp > Disp Mode > Threshold
Limit State
:CHSCanner:DISPlay:LIMit[:STATe] OFF|ON|0|1
:CHSCanner:DISPlay:LIMit[:STATe]?
Use this command to turn on/off limit state in channel
scanner mode.
Example:
:CHSCanner:DISPlay:LIMit ON
Key access:
Shift > Limit > Limit
Limit Value
:CHSCanner:DISPlay:LIMit:VALue <value>
:CHSCanner:DISPlay:LIMit:VALue?
Specifies the limit value in channel scanner mode.
Example:
:CHSCanner:DISPlay:LIMit:VALue -20
Key access:
Shift > Limit > Limit