Command Reference
Reflection Measurement Option Subsystem
140
Reflection Measurement Trace Math Type
[:SENSe]:CAT:TRACe:MATH OFF|ADD|SUB|DIV
[:SENSe]:CAT:TRACe:MATH?
This command toggles the trace math type of reflection
measurement. RM7 option support three trace type,
measurement trace data plus memory saved trace data,
immurement trace data minus memory saved trace data, and
memory saved data minus measurement trace data.
Example:
:CAT:TRACe:MATH ADD
Key access:
Trace > Trace Math
Reflection Measurement Trace Average Number
[:SENSe]:CAT:AVERage:COUNt <value>
[:SENSe]:CAT:AVERage:COUNt?
This command defines the trace average number of CAT test.
[:SENSe]:CAT:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:CAT:AVERage[:STATe]?
Use this command to turn on/off the trace average
operation.
Example:
:CAT:AVERage:COUNt 20
Key access:
Trace > Average Number
Reflection Measurement Trace Data Format
:FORMat[:TRACe][:DATA] ASCii|REAL
:FORMat[:TRACe][:DATA]?
This command toggles the trace data format. This command
affects only return trace data format. ASC stands for ASCII
string be separated by comma, and REAL stands for 4 bytes
length float without comma.
Range:
ASCii, REAL
Example:
:FORMat ASC