Operation and Features
2
Programming the Overvoltage Protection
E3632A User’s Guide
37
1
Readjust the OVP trip level or the output voltage level.
Lower the output voltage level below the OVP trip point
after pressing
or raise the OVP trip level
by using the knob after pressing
.
2
Move to the clear mode.
OVP ON
You will see the above message after pressing
. If you changed the output voltage level,
press the
twice. Turn the knob to the right
until the
OVP CLEAR
message appears on the display.
3
Clear the overvoltage condition and exit this menu.
Now, when you press
again, the
DONE
message is displayed for a second and the
OVP
annunciator will not flash any more. The output will
return to meter mode.
Remote interface operation
or
Over Voltage
Display Limit
Display Limit
Over Voltage
Over Voltage
Over Voltage
Over Voltage
Over Voltage
Over Voltage
VOLT:PROT {<voltage>|MIN|MAX}
Set the OVP level.
VOLT:PROT:STAT {OFF|ON)
Disable or enable the
OVP circuit.
VOLT:PROT:CLE
Clear the tripped
OVP circuit.
N O T E
The power supply’s OVP circuit contains a crowbar SCR, which effectively
shorts the output of the power supply whenever the overvoltage condition
occurs. If external voltage source such as a battery is connected across
the output, and the overvoltage condition inadvertently occurs, the SCR
will continuously sink a large current from the source and possibly damage
the power supply. To avoid this a diode must be connected in series with
the output as shown in
.
Summary of Contents for E3632A
Page 1: ...Keysight E3632A DC Power Supply User s Guide ...
Page 2: ......
Page 3: ......
Page 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...