4
Error Messages
Calibration Error Messages
138
E3632A User’s Guide
715
Bad OCP cal data
The overcurrent protection calibration constant is out of range. Note that the new calibration constants
are not stored in the non-volatile memory.
716
Bad DAC DNL error correction data
Invalid data measured during the calibration for DAC differential nonlinearity error correction.
717
Cal OVP or OCP status enabled
Overvoltage protection status or overcurrent protection status is enabled. You must set both overvoltage
and overcurrent protection status to OFF before and during the calibration.
740
Cal checksum failed, secure state
741
Cal checksum failed, string data
742
Cal checksum failed, store/recall data in location 0
743
Cal checksum failed, store/recall data in location 1
744
Cal checksum failed, store/recall data in location 2
745
Cal checksum failed, store/recall data in location 3
746
Cal checksum failed, DAC cal constants
747
Cal checksum failed, readback cal constants
748
Cal checksum failed, GPIB address
749
Cal checksum failed, internal data
750
Cal checksum failed, DAC DNL error correction data
Table 4-3
Calibration error messages (continued)