Error Messages
4
Execution Error Messages
E3632A User’s Guide
131
-113
Undefined header
A command was received that is not valid for this power supply. You may have misspelled the command
or it may not be a valid command. If you are using the short form of the command, remember that it may
contain up to four letters.
Example:
TRIGG:DEL 3
-114
[1]
Header suffix out of range
The numeric suffix attached to a command header is not one of the allowable values.
Example:
OUTP2 ON
-120
[1]
Numeric data error
An invalid number was specified for a numeric parameter.
Example:
APPL 1.0E+320000
-121
Invalid character in number
An invalid character was found in the number specified for a parameter value.
Example:
*ESE #B01010102
-123
Numeric overflow
A numeric parameter was found whose exponent was larger than 32000.
-124
Too many digits
A numeric parameter was found whose mantissa contained more than 255 digits, excluding leading
zeros.
-128
Numeric data not allowed
A numeric parameter was received but a character string was expected.
Example:
DISP:TEXT 123
-130
[1]
Suffix error
A suffix was incorrectly specified for a numeric parameter. You may have misspelled the suffix or the
numeric parameter does not accept a suffix.
Example:
TRIG:DEL 0.5 SECS
-131
Invalid suffix
A suffix was incorrectly specified for a numeric parameter. You may have misspelled the suffix.
Example:
TRIG:DEL 0.5 SECS
-134
Suffix too long
A suffix for a numeric parameter contained too many characters.
Table 4-1
Execution error messages (continued)
Summary of Contents for E3632A
Page 1: ...Keysight E3632A DC Power Supply User s Guide ...
Page 2: ......
Page 3: ......
Page 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...