Characteristics and Specifications
7
Supplemental Characteristics
E3632A User’s Guide
171
Supplemental Characteristics
Table 7-3
Supplemental characteristics
Output programming range
(maximum programmable values)
Low range
0 V to 15.45 V
0 V to 7.21 A
High range
0 V to
20.6 V
0 V to 1.545 A
OVP
1 V to 32 V
OCP
0 A to 7.5 A
Remote sensing capability
Voltage drop
Up to 1 V per each lead
Load regulation
Add 5 mV to specification for each 1 V change in the
output
lead due to load current changes.
Load voltage
Subtract the voltage drop in the load leads from the specified
output voltage rating.
Temperature coefficient
[1]
(% of output
offset)
Voltage
0.01%
3 mV
Current
0.02%
3 mA
Stability
[2]
(% of output
offset)
Voltage
0.02%
1 mV
Current
0.1%
1 mA
Voltage programming speed
[3]
Up
•
50 ms (full load)
•
20 ms (no load)
Down
•
45 ms (full load)
•
400 ms (no load)
Output terminal isolation
(maximum, from chassis ground)
•
60 Vdc when connecting shorting conductors without
insulation between the (
) output and the (
) sense
terminals and between the (–) output and the (–) sense
terminals.
•
240 Vdc when connecting insulated shorting conductors
between the (
) output and the (
) sense terminals and
between the (–) output and the (–) sense terminals.
AC input ratings
(selectable via rear panel selector)
•
115 Vac
10%, 47 Hz to 63 Hz (standard)
•
230 Vac
10%. 47 Hz to 63 Hz (option 0E3)
•
100 Vac ±10%, 47 Hz to 63 Hz (option 0E9)
Summary of Contents for E3632A
Page 1: ...Keysight E3632A DC Power Supply User s Guide ...
Page 2: ......
Page 3: ......
Page 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...