Remote Interface Reference
3
Output Settings and Operation Commands
E3632A User’s Guide
87
CURRent:TRIGgered
{<current>|MINimum|MAXimum}
This command programs the pending triggered current level.
The pending triggered current level is a stored value that is
transferred to the output terminals when a trigger occurs. A
pending triggered level is not affected by subsequent
CURRent
commands.
CURRent:TRIGgered? [MINimum|MAXimum]
This query returns the triggered current level presently
programmed. If no triggered level is programmed, the
CURRent
level is returned.
CURR:TRIG? MAX
and
CURR:TRIG? MIN
return the highest and lowest
programmable triggered current levels.
CURRent:PROTection
{<current>|MINimum|MAXimum}
This command sets the current level at which the
overcurrent protection (OCP) circuit will trip. If the peak
output current exceeds the OCP level, then the output
current is programmed to zero. The Questionable Status
register “OC” bit is set (refer to
). An overcurrent condition can be cleared with the
CURR:PROT:CLE
command after the condition that caused
the OCP trip is removed.
CURRent:PROTection? {MINimum|MAXimum}
This query returns the overcurrent protection trip level
presently programmed.
CURR:PROT? MAX
and
CURR:PROT?
MIN
return the maximum and minimum programmable
overcurrent trip levels.
Summary of Contents for E3632A
Page 1: ...Keysight E3632A DC Power Supply User s Guide ...
Page 2: ......
Page 3: ......
Page 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...