6
Tutorial
Connecting the Load
158
E3632A User’s Guide
Remote voltage sensing
Normally, a power supply operating in the constant voltage
mode achieves its optimum line and load regulations, its
lowest output impedance, drift, and ripple and noise, and its
fastest transient recovery performance at the power supply
output terminals. If the load is separated from the output
terminals by any lead length, some of these performance
characteristics will be degraded at the load terminals —
usually by an amount proportional to the impedance of the
load leads compared with the output impedance of the
power supply.
Table 6-1
Wire rating
AWG
10
12
14
16
18
20
22
24
26
28
Suggested
maximum
Current(amps)
[1]
40
25
20
13
10
7
5
3.5
2.5
1.7
m
ft
1.00
1.59
2.53
4.02
6.39
10.2
16.1
25.7
40.8
64.9
m
m
3.3
5.2
8.3
13.2
21.0
33.5
52.8
84.3
133.9
212.9
[1] Single conductor in gree air at 30 °C with insulation
WA R N I N G
To satisfy safety requirements, load wires must be heavy enough not
to overheat while carrying the short-circuit output current of the
power supply.