Operation and Features
2
System-Related Operations
E3632A User’s Guide
47
Error conditions
When the front- panel
ERROR
annunciator turns on, one or
more command syntax or hardware errors have been
detected. A record of up to 20 errors can be stored in the
power supply’s error queue. Refer to
for more information.
•
Errors are retrieved in first- in- first- out (FIFO) order. The
first error returned is the first error that was stored.
When you have read all errors from the queue, the
ERROR annunciator turns off. The power supply beeps
once each time an error is generated.
•
If more than 20 errors have occurred when you operate
the power supply over the remote interface, the last error
stored in the queue (the most recent error) is replaced
with - 350, “Queue overflow”. No additional errors are
stored until you remove errors from the queue. If no
errors have occurred when you read the error queue, the
power supply responds with +0, “No error” over the
remote interface or
NO ERRORS
from the front panel.
•
The error queue is cleared when power has been off or
after a
*CLS
(clear status) command has been executed.
The
*RST
(reset) command does not clear the error
queue.
Front panel operation
If the
ERROR
annunciator is on, press
repeatedly to
read the errors stored in the queue. All errors are cleared
when you read all errors.
ERROR -113
Remote interface operation
Errors have the following format (the error string may
contain up to 80 characters).
- 113, “Undefined header”
SYST:ERR?
Reads an error from the error queue
Error
Summary of Contents for E3632A
Page 1: ...Keysight E3632A DC Power Supply User s Guide ...
Page 2: ......
Page 3: ......
Page 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...