4
Error Messages
Execution Error Messages
134
E3632A User’s Guide
-440
Query UNTERMINATED after indefinite response
The
*IDN?
command must be the last query command within a command string.
Example:
*IDN?;:SYST:VERS?
501
Isolator UART framing error
502
Isolator UART overrun error
503
[1]
SPI data error
Data error was detected during the communication between the main controller U10 and the I/O
controller U752.
511
RS-232 framing error
512
RS-232 overrun error
513
RS-232 parity error
514
Command allowed only with RS-232
There are three commands which are only allowed with the RS-232 interface:
SYSTem:LOCal
,
SYSTem:REMote
, and
SYSTem:RWLock
.
521
Input buffer overflow
522
Output buffer overflow
550
Command not allowed in local
You should always execute the
SYSTem:REMote
command before sending other commands over the
RS-232 interface.
[1] This error message is only applicable for serial MY53xx6xxx.
Table 4-1
Execution error messages (continued)