3
Remote Interface Reference
Status Reporting Commands
118
E3632A User’s Guide
STATus:QUEStionable:CONDition?
This command queries the Questionable Status condition
register to check CV or CC mode of the power supply. The
power supply returns a decimal value which corresponds to
the binary- weighted sum of all bits in the register. These
bits are not latched. If
0
is returned, the power supply is in
output off or unregulated state. If
1
is returned, the power
supply is in the CC operating mode and if
2
is returned, the
power supply is in the CV operating mode. If
3
is returned,
the power supply is in failure.
STATus:QUEStionable?
This command queries the Questionable Status event
register. The power supply returns a decimal value which
corresponds to the binary- weighted sum of all bits in the
register. These bits are latched. Reading the event register
clears it.
STATus:QUEStionable:ENABle <enable value>
This command enables bits in the Questionable Status
Enable register. The selected bits are then reported to the
Status Byte.
STATus:QUEStionable:ENABle?
This command queries the Questionable Status Enable
register. The power supply returns a binary- weighted
decimal representing the bits set in the enable register.
*CLS
This command clears all event registers and Status Byte
register.
*ESE <enable value>
This command enables bits in the Standard Event Enable
register. The selected bits are then reported to the Status
Byte.
Summary of Contents for E3632A
Page 1: ...Keysight E3632A DC Power Supply User s Guide ...
Page 2: ......
Page 3: ......
Page 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...