54
Chapter 4 Calibration Procedures
Time Required for Calibration
4
Time Required for Calibration
The
Keysight
33210A can be automatically calibrated under computer
control. With computer control you can perform the complete calibration
procedure and performance verification tests in approximately 30 minutes
once the instrument is warmed-up (see “Test Considerations” on page 57).
Manual adjustments and verifications, using the recommended test
equipment, will take approximately 2 hours.
START
Incoming
Verification?
YES
NO
Perform
Adjustments
(approx 1 Hour)
Do Performance
Verification Tests
(approx 1 Hour)
Do Performance
Verification Tests
(approx 1 Hour)
DONE