123
Chapter 6 Disassembly and Repair
Mechanical Disassembly
4
6
External Timebase Circuit Board Disassembly
The External Timebase is an option to the basic instrument (Option 001).
Remove the hex nuts and washers holding the rear panel
10 MHz In
and
10 MHz Out
BNC connectors. Pull the External Timebase PC board (A3
assembly) from the back panel as shown below. Then unplug the captive
ribbon cable from the keyed connector on the main PC board (J601 on the
A1 assembly). Remove the A3 assembly from the chassis.
A3