88
Chapter 4 Calibration Procedures
+10 dB Range Flatness Adjustments
4
* Constants are stored after completing this setup.
3
Using the numeric keypad, adjust the displayed amplitude at each setup
to match the measured amplitude (in dBm). Then select
ENTER VALUE
.
Note
In order to get dBm you must use the numeric keypad (not the knob) to
enter the number, and then select “dBm”.
4
After performing setup 62:
a. If your calibration procedures require you to verify the adjustment
just made, exit the calibration menu and perform “+10 dB Range
Flatness Verification”, on page 67.
b. If you are making all the adjustments and then verifying the
instrument’s performance, continue with the next procedure in
this chapter.
Nominal Signal
Setup
Frequency
Amplitude
54*
100 kHz
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
55*
200 kHz
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
56*
500 kHz
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
57*
1.5 MHz
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
58*
3.0 MHz
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
59
4.0 MHz
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
60*
6.0 MHz
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
61*
8.0 MHz
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
62*
9.9 MHz
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator Service Guide ...
Page 2: ......
Page 3: ......
Page 10: ...8 ...
Page 13: ...1 1 Specifications ...
Page 19: ...2 2 Quick Start ...
Page 37: ...3 3 Front Panel Menu Operation ...
Page 52: ...50 3 ...
Page 53: ...4 4 Calibration Procedures ...
Page 96: ...94 Chapter 4 Calibration Procedures Calibration Errors 4 ...
Page 97: ...5 5 Block Diagram ...
Page 101: ...99 Chapter 5 Block Diagram Block Diagram 4 5 ...
Page 103: ...6 6 Disassembly and Repair ...
Page 128: ...126 Chapter 6 Disassembly and Repair Replaceable Parts 6 ...
Page 129: ...7 7 Backdating ...