82
Chapter 4 Calibration Procedures
AC Amplitude (high-impedance) Adjustment
4
2
Use the DMM to measure the dc voltage at the front-panel Output
connector for each setup in the following table.
* Constants are stored after completing this setup.
Nominal Signal
Setup
DC level
13
+0.015 V
Output of -30dB range
14*
-0.015 V
Output of -30dB range
15
+0.05 V
Output of -20dB range
16*
-0.05 V
Output of -20dB range
17
+0.15 V
Output of -10dB range
18*
-0.15 V
Output of -10dB range
19
+0.50 V
Output of 0dB range
20*
-0.50 V
Output of 0dB range
21
+0.15 V
Output of -10dB range (Amplifier In)
22*
-0.15 V
Output of -10dB range (Amplifier In)
23
+0.50 V
Output of 0dB range (Amplifier In)
24*
-0.50 V
Output of 0dB range (Amplifier In)
25
+1.5 V
Output of +10dB range (Amplifier In)
26*
-1.5 V
Output of +10dB range (Amplifier In)
27
+5 V
Output of +20dB range (Amplifier In)
28*
-5 V
Output of +20dB range (Amplifier In)
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator Service Guide ...
Page 2: ......
Page 3: ......
Page 10: ...8 ...
Page 13: ...1 1 Specifications ...
Page 19: ...2 2 Quick Start ...
Page 37: ...3 3 Front Panel Menu Operation ...
Page 52: ...50 3 ...
Page 53: ...4 4 Calibration Procedures ...
Page 96: ...94 Chapter 4 Calibration Procedures Calibration Errors 4 ...
Page 97: ...5 5 Block Diagram ...
Page 101: ...99 Chapter 5 Block Diagram Block Diagram 4 5 ...
Page 103: ...6 6 Disassembly and Repair ...
Page 128: ...126 Chapter 6 Disassembly and Repair Replaceable Parts 6 ...
Page 129: ...7 7 Backdating ...