28
Chapter 2 Quick Start
To Configure a Pulse Waveform
2
To Configure a Pulse Waveform
You can configure the function generator to output a pulse waveform
with variable pulse width and edge time. The following steps show you
how to configure a 500 ms pulse waveform with a pulse width of 10 ms
and edge times of 50 ns.
1 Select the pulse function.
Press the
key to select the pulse function and output a pulse
waveform with the default parameters.
2 Set the pulse period.
Press the
Period
softkey and then set the pulse period to 500 ms.
3 Set the pulse width.
Press the
Width
softkey and then set the pulse width to 10 ms. The pulse
width represents the time from the 50% threshold of the rising edge to
the 50% threshold of the next falling edge (note the display icon).
4 Set the edge time for both edges.
Press the
Edge Time
softkey and then set the edge time for both the
rising and falling edges to 50 ns. The edge time represents the time from
the 10% threshold to the 90% threshold of each edge (note the display icon).
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator Service Guide ...
Page 2: ......
Page 3: ......
Page 10: ...8 ...
Page 13: ...1 1 Specifications ...
Page 19: ...2 2 Quick Start ...
Page 37: ...3 3 Front Panel Menu Operation ...
Page 52: ...50 3 ...
Page 53: ...4 4 Calibration Procedures ...
Page 96: ...94 Chapter 4 Calibration Procedures Calibration Errors 4 ...
Page 97: ...5 5 Block Diagram ...
Page 101: ...99 Chapter 5 Block Diagram Block Diagram 4 5 ...
Page 103: ...6 6 Disassembly and Repair ...
Page 128: ...126 Chapter 6 Disassembly and Repair Replaceable Parts 6 ...
Page 129: ...7 7 Backdating ...