80
Chapter 4 Calibration Procedures
Output Impedance Adjustment
4
2
Use the DMM to make a resistance measurement at the front panel
Output connector for each setup in the following table. The expected
measured value is approximately 50
Ω
.
* Constants are stored after completing this setup.
3
Using the numeric keypad or knob, adjust the displayed impedance at
each setup to match the measured impedance. Select
ENTER VALUE
.
4
There are no specific operational verification tests for Output
Impedance. Continue with the next adjustment procedure in this
chapter.
Setup
8*
-30dB range
9*
-20dB range
10*
-10dB range
11*
0dB range
12*
+10dB range