76
Chapter 4 Calibration Procedures
Self-Test
4
Self-Test
Self-Test is performed as the first step to ensure the instrument is in
working order before beginning any additional adjustments.
Note
Be sure to follow the requirements listed in “Test Considerations” on
page 57 before beginning any adjustments.
1
Press
on the front panel. Select
Perform Cal
on the “Test / Cal”
menu. Enter setup number “1” and select
BEGIN
.
2
If the instrument fails any self-test, you must repair the instrument
before continuing the adjustment procedures.
Note
The self-test procedure takes approximately 15 seconds to complete.
Setup
1
Performs the Self-test. The Main Output is disabled during test.