Keysight U3042AM20 User’s and Service Guide
77
U3042AM20
Troubleshooting the Test Set
RF Switching Failures
If the test set’s RF signal functionality is not operating properly, use the
following procedure to aid in isolating the source of problem. The RF internal
mechanical switches used in this instrument are biased with the +24 Vdc
supply voltage and controlled logic signals coming from the test set Controller
board and Switch Interface board.
1. Place the network analyzer and test set side by side (not stacked) and
connect the I/O cable (N4011-21002). See
2. Ensure that the network analyzer is not in Multiport Mode.
3. Using the Test Set I/O Data command values, confirm the correct address
and data values are used, refer to
“Address and Data Values” on page 47
4. Verify that the ribbon cable is installed on the correct connector on the
Switch Interface board. If the connection is correct, substitute a working
switch or ribbon cable. Swapping can be performed in this situation.
Test Set and RF Switching Path Test
If you suspect an RF signal path problem with the test set and have verified
that the problem in not the front panel RF interface cables, the following
procedure will check all of the RF signal paths through the test set.
— Install the test set I/O cable from the analyzer to the Test Set rear panel.
— Remove the front panel RF interface cables and reinstall the analyzer’s front
panel jumper cables.
Equipment Required
— PNA, PNA-L or PNA-X Network Analyzer
— Two RF Flex Cables (3.5 mm male)
— 3.5 mm Adapters (female to female)
— Test Set I/O Cable