Keysight U3042AM20 User’s and Service Guide
59
U3042AM20
Cal Kit Operational Check
Cal Kit Operational Check
The following procedure can be used to confirm that the test set and network
analyzer are operational. The operation verification limits provided ensure that
your test set and network analyzer is operating properly. Refer to
Verification Limits
Specifications for the Test Set are typical. System performance for the analyzer
and test set are only characteristic and intended as non-warranted
information. Only a functional certificate is provided for the Test Set.
It is recommended that you return your instrument to Keysight Technologies
for servicing or repair if the Test Set and PNA-X performance exceed the
operational verification limits.
A periodic calibration is not required. The Operators Check should be
performed after System Setup, or if performance is in question. An N-Port
calibration should be performed before making a measurement.
If you suspect that your 24-Port configuration is not operating properly,
ensure that all front RF jumper interconnect cables are correctly attached.
Table 16
N5242A Operation Verification Limits for Reflection Tracking
1
1. Reflection Tracking takes into account Source Loss, Receiver Loss,
Margin, and PNA-X Mixer Cal.
Frequency
Port 1-24
10 MHz to 4 GHz
−
8 dB
4 GHz to 6 GHz
−
9 dB
6 GHz to 10 GHz
−
10 dB
10 GHz to 15 GHz
−12
dB
15 GHz to 20 GHz
−
14 dB
20 GHz to 26.5 GHz
−
15 dB
Summary of Contents for U3042AM20