Keysight U3042AM20 User’s and Service Guide
7
U3042AM20
General Specifications
Environmental Requirements
Refer to your network analyzer’s standard documentation for environmental
requirements.
The test set complies with all applicable safety and regulatory requirements for
the intended location
of use.
— Pressure Altitude (Operation)
3,000 meters (~10,000 feet)
— The instrument can safely operate in a relative humidity of 80% for
temperatures to 31 degrees C, decreasing linearly to 50% relative humidity
at 40 degrees C.
Equipment Heating and Cooling
If necessary, install air conditioning and heating to maintain the ambient
temperature within the appropriate range.
Ventilation Requirements: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4
°
C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
forced convection must be used.
Required Conditions for Accuracy Enhanced Measurement
Accuracy enhanced (error–corrected) measurements require the ambient
temperature of the analyzer and test set to be maintained within ± 1 °C of the
ambient temperature at calibration.
Dimensions and Space Requirements
Standard installation of the test set and analyzer includes configuration and
installation on a customer provided lab bench or table top of adequate size and
strength. For weight, dimensions and space requirements, refer to the network
analyzer documentation that is used to configure the test set.