Keysight U3042AM20 User’s and Service Guide
61
U3042AM20
Cal Kit Operational Check
10. Select Stimulus > Sweep > Number of Points > 401.
11. Connect the ECal module to an available PNA-X USB port on the front or
. This procedure assumes you are
using a ECal. If you are not, see
“1-Port Calibration and Verification
See “Perform a 1-Port Calibration on Port 1. On
the analyzer, select Response > Cal > Start Cal > Calibration Wizard.” on
page 62.
12. Allow the ECal module, Test Set and PNA-X to warm up for a minimum of
30 minutes.